Relay Assignments
The 9125 controller has the option of using eight relay outputs. All of the relays have a positive common terminal and independent negative terminals. All of the relays are configured in a normally closed position except relay number eight, which has both a normally closed (NC) and a normally open (NO) terminal. These relays can be configured to work with events, alarms, loops, burnoff and alarm combinations.
Relay Output Terminals
Relay Output 1 – terminals 7 and 8
Relay Output 2 – terminals 7 and 9
Relay Output 3 – terminals 7 and 10
Relay Output 4 – terminals 7 and 11
Relay Output 5 – terminals 7 and 12
Relay Output 6 – terminals 7 and 13
Relay Output 7 – terminals 7 and 14
Relay Output 8 – terminals 7 and 15 NC
Relay Output 8 – terminals 7 and 16 NO
Relay Output Choices
Loop 1 Fwd Loop 1 Rev Loop 2 Fwd Loop 2 Rev Loop 3 Fwd Loop 3 Rev Alarm 1 - 3
|
Event 0 - 15 Burnoff IN1 Relay SP A IN1 Relay SP B IN1 Relay SP C IN2 Relay SP A IN2 Relay SP B
|
IN2 Relay SP C IN3 Relay SP A IN3 Relay SP B IN3 Relay SP C IR Sample Solenoid Alarm Combination N/A
|
The “Alarm Combination” option will allow the user to select the specific combination of alarms to use.
The options are: Alarm 1, Alarm 2, Alarm 3, Invert Relay State for Alarms, and one of two options: T/C Check (Thermocouple Check), or EOQ (End of Quench)..